Clinical Transportable Non Vacuum Steam Steriliser: Automatic Control Test

Testing Clinical Transportable Non Vacuum Steam Sterilisers

Automatic Control Test

This test is performed to demonstrate that the sterilisation cycle is working within validated parameters such as time, temperature and pressure, using the machines own indicated values on the display.

This test should only be undertaken by a competent and trained person.

The test is required to be completed for each cycle type that the steriliser operates.

Document Specific Test Schedule

Factory
Testing
Installation
& Commissioning
Periodic
Testing
Document Specific Title Reference TT WT IQ OQ PQ D W Q Y
HTM 01-05 Automatic Control Test 15.4          
HTM 2010
Part 3: Validation & Verification
Automatic Control Test 12.1       
BS EN 13060
TT/WT/IQ
BS EN ISO 17665
OQ/PQ/DWQY
Automatic Control Test
10.3 (C)
BS EN ISO
17665:1