Temperature Calibrators

Temperature Calibrators

For use in performing pre and post calibration checks of measurement sensors, we provide popular temperature calibrators from industry leading manufacturers such as Ametek and Isothermal Technologies.

See our Comparison Guide for Your Temperature Calibrator at the bottom of the page >


 

JOFRA ITC 155A
Temperature Calibrator

Range:
-23°C to 155°C
Removable Insert:
8x 3.0mm
1x 1/4" Reference

Complete With Traceable Calibration Certificate and Robust Alluminium Carry Case.

One of our most popular temperature calibrators. Isopharm advise that the internal reference probe is within the accuracy budget required by standards. You can therefore still take advantage of the TQSoft automatic calibration feature using the unit's internal reference probe.

A perfect solution for use in performing pre and post calibration checks when validating instrument washer disinfector (IWD) and steriliser processes.

See this product in our online shop >

Get your own validation system quotation here >


Fastcal Complete
Temperature Calibrator

Range:
-35°C to +140°C
Insert:
2x 4.5mm, 1x 6.5mm, 1x 8.0mm

Complete With Independent Reference ProbeTraceable Calibration Certificate and Soft Carry Case.

One of our most popular temperature calibrators. RS232 and the in-built independent reference probe allows complete automatic temperature calibration with TQSoft.

A perfect solution for use in performing pre and post calibration checks when validating endoscope washer disinfector (EWD), instrument washer disinfector (IWD) and steriliser processes.

See this product in our online shop >

Get your own validation system quotation here >


JOFRA CTC 320A
Temperature Calibrator

Range:
+33°C to +320°C
Insert:
1x 3.0mm, 1x 4.0mm, 1x 5.0mm, 1x 6.0mm, 1x 9.0mm 

Complete With Traceable Calibration Certificate and Robust Alluminium Carry Case.

One of our most popular temperature calibrators. RS232 allows automatic calibration with TQSoft when used with a reference probe such as the Ametek DTI050.

A perfect solution for use in performing pre and post calibration checks when validating instrument washer disinfector (IWD) and steriliser processes.

See this product in our online shop >

Get your own validation system quotation here >


JOFRA CTC 140A
Temperature Calibrator

Range:
-17°C to +140°C
Insert:
1x 3.0mm, 1x 4.0mm, 1x 5.0mm, 1x 6.0mm

Complete With Traceable Calibration Certificate and Robust Alluminium Carry Case.

RS232 allows automatic calibration with TQSoft when used with a reference probe such as the Ametek DTI050.

A perfect solution for use in performing pre and post calibration checks when validating endoscope washer disinfector (EWD), instrument washer disinfector (IWD) and steriliser processes.

See this product in our online shop >

Get your own validation system quotation here >


JOFRA ETC 400A
Temperature Calibrator

Range:
+28°C to +400°C
Insert:
1x 2.0mm, 1x 3.0mm, 2x 4.0mm, 1x 6.0mm

Complete With Traceable Calibration Certificate.

RS232 allows automatic calibration with TQSoft when used with a reference probe such as the Ametek DTI050.

A perfect solution for use in performing pre and post calibration checks when validating instrument washer disinfector (IWD) or steriliser processes.

See this product in our online shop >

Get your own validation system quotation here >


 

A Comparison Guide for Your Temperature Calibrator

 

         
  FASTCAL COMPLETE JOFRA
CTC 320A
JOFRA
CTC 140A
JOFRA
ETC 400A
JOFRA
ITC 155A
Range -35°C to +140°C +33°C to +320°C -17°C to +140°C +28°C to +400°C -23°C to 155°C
Stability 0.02°C ±0.1°C ±0.05°C ±0.15°C ±0.01°C
Reference Probe YES 1 NO 2 NO 2 NO 2 NO 4
Accuracy with External Reference Probe ±0.15°C N/A N/A N/A N/A
Accuracy without External  Reference Probe ±0.2°C ±0.4°C 3 ±0.5°C 3 ±0.5°C 3 ±0.18°C
TQSoft Automatic Calibration YES YES YES YES YES
Heating Time (Approximately) Ambient to +140°C
10 minutes
Ambient to +140°C
5 minutes
Ambient to +140°C
10 minutes 
Ambient to +140°C
4 minutes
Ambient to +140°C
10 Minutes
Well Sizes 2x 4.5mm
1x 6.5mm
1x 8.0mm
1x 3.0mm
1x 4.0mm
1x 5.0mm
1x 6.0mm
1x 9.0mm
1x 3.0mm
1x 4.0mm
1x 5.0mm
1x 6.0mm
1x 2.0mm
1x 3.0mm
2x 4.0mm
1x 6.0mm
Standard
8x 3.0mm
1x 1/4" Reference
(Removable Insert,
see options above.)
Physical Dimensions
L x W x H
228 x 248 x 143mm 241 x 139 x 325mm 241 x 139 x 325mm 172 × 72 × 182mm 241 x 139 x 375mm
Certificate Provided Traceable Traceable Traceable Traceable Traceable
Carry Case Provided Soft Case Aluminium Case Aluminium Case None Aluminium Case
1: This unit has an in-built independent reference probe so can take advantage of the TQSoft automatic calibration feature.
2: Accuracy can be greatly improved when using a compatible reference probe and the TQSoft automatic calibration feature.
3: This unit will require an independent reference probe in order to take advantage of the TQSoft automatic calibration feature.
4: Isopharm advise that the ITC-155's internal control sensor is accurate enough to ensure your system is still within the accuracy budget required by standards without the use of a reference probe. You can therefore still take advantage of the TQSoft automatic calibration feature using the unit's internal probe.