Ultrasonic Cleaner: Cleaning Efficacy Soil Test - Chamber/Load Carrier/Reference Load

Testing Ultrasonic Cleaners

Cleaning Efficacy Soil Test - Chamber/Load Carrier/Reference Load

This test is performed to show that the ultrasonic cleaner can successfully process a ‘worse case’ load.

This test should only be undertaken by a competent and trained person.

The test is required to be completed for each cycle type that the ultrasonic cleaner operates.

Document Specific Test Schedule

Factory
Testing
Installation
& Commissioning
Periodic
Testing
Document Specific Title Reference TT WT IQ OQ PQ D W Q Y
CFPP
01-01
Part D

Cleaning Efficacy Test – Test Soil

Cleaning Efficacy Test – Chamber Walls & Load Carriers*

Cleaning Efficacy Test – Reference Load*

2.240

2.242

2.247

       
NI/CFPP/02

Cleaning Efficacy Test – Chamber Walls & Load Carriers*

2.242
CFPP 01-01 Part D

       

HTM 01-05

Cleaning Efficacy Test

14.3
Refers to ISO 15883

       

HTM 2030
Validation & Verification

Cleaning Efficacy Test – Reference Load*

Cleaning Efficacy Test – Test Soil*

9.18

 

19.22

       

SHTM 2030
Validation & Verification

Cleaning Efficacy Test – Chamber Walls & Load Carriers

6.28

       

WHTM
01-01
Part D

Cleaning Efficacy with Test Soil (Chamber Walls & Load Carriers)*

Cleaning Efficacy with Test Soil (Reference Load)*

2.237

 

2.242

       
WHTM 01-05
Cleaning Efficacy Test

14.3
Refers to ISO 15883

           
BS EN ISO 15883
Part 1

Cleaning Efficacy Test 1 – Chamber

Cleaning Efficacy Test 1 – Load Carrier

Cleaning Efficacy Test 1 – Load

6.10.2          

* Cleaning efficacy test on Reference Load is performed directly after a successful cleaning efficacy test on Walls and Load Carriers - Ref: 2.247 (CFPP 01-01 Part D) & 10.20 (HTM 2030)